The EECPC 3790A-10 fully meets all PMP requirements of upcoming post-Euro 6 regulations for light and heavy-duty vehicles, going down to 10nm. It builds upon the proven rugged, reliable, and highly repeatable performance of TSI's butanol CPC technology. The full-flow design incorporates new improvement and performance features, such as anti-spill/anti-flooding design, condensate removal system, removable saturator for easy maintenance, and an internal pulse height monitor to ensure accurate performance. The EECPC 3790A-10 has a lower particle size detection limit (D50) of 10nm.
Applications
- Diesel, gasoline and CNG engine emissions testing
- Aircraft turbine eengine solid particle emissions testing
Included Items
- Aerosol Instrument Manager® software (CPC module)
Features and benefits
- Meets the PMP lower particle size detection limit at the defined D50 and D90 particle sizes
- Achieves a linear response to particle concentration from 1 to >50,000 particles/ cm3 with R2 >0.97 with live-time coincidence correction
- False background concentration below 0.001 particles/cm3
- Achieves a counting accuracy of ± 10% against a traceable standard
- Operates under full flow conditions using single particle counting
- Calibration is in full compliance with PMP requirements and ISO 27891
- Achieves readability of 0.1 particles/cm3
- Internal pulse-height monitor to ensure accurate performance
- 10 Hz data rate for modal analysis